Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 119
Alarms
Cardiac bypass mode
Cardiac bypass mode is only available when the
Cockpit is in OR mode. When cardiac bypass mode
is activated (see page 460), the following happens
at the Cockpit:
– All alarm monitoring (including arrhythmia
alarms), and the alarm bar are deactivated.
– Alarm reminders are not activated.
– The message All alarms off: bypass and the
symbol appear in the message area in the
upper-right corner of the header (in white text
on red background). This message also
displays at the ICS and on a remote device.
– Pressing the All alarms paused or the All
alarms off %0 button deactivates cardiac
bypass mode.
– Pressing the yellow key on the Cockpit does not
pause any alarms.
– The non-invasive blood pressure interval mode
is deactivated. The interval timer is restored to
the last value. To restart the interval
measurements, press the NIBP start/stop
button or the start/stop key.
– The alarm history records either the message
Cardiac bypass on or Cardiac bypass off.
When the Cockpit is in cardiac bypass mode and
the M540 is undocked, cardiac bypass mode is not
supported on the M540. Cardiac bypass is
activated on the M540 after it docks to a Cockpit
that is in cardiac bypass mode.
To activate or deactivate this function, see page
460. If French NFC mode is activated, Cardiac
bypass mode is not available.
OR alarms
When OR alarms are activated, alarm messages
for medium and high-priority alarms clear when the
alarm condition no longer exists. In addition, you
can deactivate the acoustic alarm signal. For
detailed information, see page 114.
French NFC mode
When this mode is activated, the following happens
at the Cockpit:
– Heart rate alarms cannot be deactivated.
– The alarm pause period cannot last longer than
3 minutes.
– You cannot activate Cardiac bypass mode
when French NFC mode is activated. If
Cardiac bypass mode was activated before
French NFC mode was activated, Cardiac
bypass mode is deactivated.
To activate or deactivate this function, see page
473.
NOTE
RRi and 12-lead ECG monitoring are unavailable
when the M540 is in OR mode and the ESU filter
is set to Monitor.