External devices – Evita 2D, Evita 4, Evita XL (Medibus)
440 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Ventilator parameter setup functions
See page 439 for information on how to access this
page.
Selection Available settings Description
Flow scale 5, 10, 15, 20 (default in neonatal mode), 35,
50, 100 (default in adult and pediatric mode),
150, 200 L/min
Determines the scale of the
displayed Flow waveform.
Vol scale 5, 10, 25, 50 (default in neonatal mode), 75,
100, 250, 500, 1000 (default in pediatric and
adult mode), 1500 mL
Determines the scale of the
displayed Volume waveform.
Parameter 1 MVe (default), RR, VTe Selects the primary parameter
in the Vent parameter field.
Parameter 2 MVe, RR (default), VTe Selects the second parameter
in the Vent parameter field.
Parameter 3 MVe, RR, VTe (default) Selects the third parameter in
the Vent parameter field.
Color Red, green, blue, yellow, light blue (default),
purple, orange, white.
Determines the color of all
ventilation parameter fields,
waveforms, and loops.