Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 521
Troubleshooting
ARR
Except for asystole and ventricular fibrillation, you
can assign the alarm priority low, medium or high or
you can deactivate the alarm function. For asystole
and ventricular fibrillation, the alarm priority is fixed
as life-threatening and you cannot deactivate the
alarm function.
!! %0 out of range
low
1)
The parameter
value is replaced
by - - -
The parameter
value is below the
measurement
range of the
monitor.
Check the patient
and treat if
necessary.
Check the placement
of electrodes and
change their position
if necessary.
!! %0 out of range
high
1)
The parameter
value is replaced
by +++
The parameter
value is above the
measurement
range of the
monitor.
Check the patient
and treat if
necessary.
Check the placement
of electrodes and
change their position
if necessary.
! %0 Unplugged
1)
The parameter
value is replaced
by ***
ECG lead wires
are disconnected
from the M540.
Check the cables and
connections; replace if
necessary.
None ST relearn The parameter
value appears
blank.
ST relearn is in
progress
Informational message –
no action required.
1)
%0 is a placeholder for the parameter label ST.
Priority Message Parameter field Problem Solution

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals