Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 547
Troubleshooting
Microstream CO2
Calibration and maintenance
Priority Message Cause Remedy
None CO
2 calibration check
failed
The Microstream MCable
calibration procedure failed.
Contact DrägerService
®
.
None CO
2 calibration check
successful
The Microstream MCable
calibration procedure was
successful.
Informational message – no
action required.
None CO
2 calibration check in
progress
The Microstream MCable
calibration procedure is in
progress.
Informational message – no
action required.
None CO
2 calibration required The Microstream MCable
calibration procedure is due.
Contact DrägerService
®
.
None CO
2 MCable:
Maintenance is due
Maintenance for
Microstream MCable is due.
Contact DrägerService
®
.
None %0 zeroing failed Resetting the Microstream
MCable to zero has failed
the standard three attempts.
Contact DrägerService
®
.
None %0 zeroing in progress The Microstream MCable is
being reset to zero.
Informational message – no
action required.
None CO2 zeroing succesful The Microstream MCable is
successfully reset to zero.
Informational message -- no
action required.