Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 135
Alarms
To change ARR alarm settings
1 Select the Alarms... button on the main menu
bar.
2 Select the Limits tab (if not already selected).
3 Select the ARR tab along the right side.
4 Select the corresponding setup button in the
Alarm priority column (C) to select the alarm
priority. A crossed-out triangle appears when
alarm monitoring is deactivated. The priority for
asystole and ventricular fibrillation events
cannot be changed. The alarm priority ’high’ is
always assigned to these categories.
5 Select the corresponding setup button in the
Rate column (D) to set the rate.
6 Select the corresponding setup button in the
Count column (E) to set the count.
7 Use one of the following settings in the Archive
column (F) to determine what happens in
response to an alarm:
– Off – no event is stored and no recording is
generated.
– Store – stores the event for later review
(see page 140).
– Record – generates a timed recording.
– Str/Rec – generates a timed recording and
stores the event.
8 Select the desired arrhythmia mode using the
Arrhythmia mode buttons (H).
Alarm setup for ST
The following diagram shows the Limits > ST page
where you configure alarm settings for ST
parameters. This page consists of a table with
setup rows for each ST parameter. Each setup row
has several fields for configuring the individual ST
alarm settings. When you select a field on the page,
an orange frame highlights the selected row to
mark your place on the setup page.