Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 135
Alarms
To change ARR alarm settings
1 Select the Alarms... button on the main menu
bar.
2 Select the Limits tab (if not already selected).
3 Select the ARR tab along the right side.
4 Select the corresponding setup button in the
Alarm priority column (C) to select the alarm
priority. A crossed-out triangle appears when
alarm monitoring is deactivated. The priority for
asystole and ventricular fibrillation events
cannot be changed. The alarm priority ’high’ is
always assigned to these categories.
5 Select the corresponding setup button in the
Rate column (D) to set the rate.
6 Select the corresponding setup button in the
Count column (E) to set the count.
7 Use one of the following settings in the Archive
column (F) to determine what happens in
response to an alarm:
Off – no event is stored and no recording is
generated.
Store – stores the event for later review
(see page 140).
Record – generates a timed recording.
Str/Recgenerates a timed recording and
stores the event.
8 Select the desired arrhythmia mode using the
Arrhythmia mode buttons (H).
Alarm setup for ST
The following diagram shows the Limits > ST page
where you configure alarm settings for ST
parameters. This page consists of a table with
setup rows for each ST parameter. Each setup row
has several fields for configuring the individual ST
alarm settings. When you select a field on the page,
an orange frame highlights the selected row to
mark your place on the setup page.

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals