Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 143
Alarms
To configure remote patient monitor alarms, see
page 145.
Configuring the SpO2 alarm priority
The following two SpO2 alarm messages can be
configured for the alarm priority that is most
appropriate for your care environment. Depending
on which MCable is used, the message reporting
the underlying alarm condition differs:
–Masimo: SpO2 sensor off
– Nellcor: SpO2 check sensor
Both alarm settings are saved as part of the patient
profile.
Configuring the alarm priority for a
Masimo sensor off message
The SpO2 sensor off message appears when the
MCable detects that the sensor is no longer
attached to the patient. The alarm priority for this
message can be configured separately for each
patient category. It is available for the following
SpO2 cables:
– Masimo rainbow SET MCable
– Masimo SET MCable
For information on how to configure the alarm
priority for the setting SpO2 sensor off, see
page 458.
Show alarm limits – On (default)
–Off
Determines whether alarm limits appear in the
parameter fields.
Cardiac bypass – On
– Off (default)
Activates/deactivates cardiac bypass mode.
Alarm functions are affected when cardiac
bypass mode is activated (see page 119).
This mode is not available when the French
NFC mode is enabled (see page 473).