EasyManuals Logo

Dräger Infinity M540 User Manual

Dräger Infinity M540
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #368 background imageLoading...
Page #368 background image
Scio Monitoring
368 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Troubleshooting
In addition to evaluating the clinical status of a
patient, CO2 waveforms can help troubleshoot
problems with equipment. The following table
shows how CO
2 waveforms can be used to identify
common problems.
Description Cause Capnogram
Alveolar plateau showing a
downward slope that merg-
es with a descending limb.
Inadequate seal around the
endotracheal tube
Leaky or deflated endotracheal
or tracheostomy cuff
Artificial airway that is too small
for the patient
Elevated waveform base-
line with corresponding in-
crease in CO
2 level.
Rebreathing due to one of the fol-
lowing causes:
Disposable airway adapter is
used although the Cockpit is
configured for the reusable
adapter type
Contaminated airway adapter
(dirty window)
–CO
2 zero drift
Insufficient expiratory time
Faulty expiratory valve
Inadequate inspiratory flow
Malfunction of a CO
2 absorber
system
Partial rebreathing circuits

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals