External devices – MEDIBUS.X devices
404 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Supported MEDIBUS.X ventilator and anesthesia data
The MEDIBUS.X ventilators and anesthesia
machines send parameters, settings, modes, and
waveforms are broadcast to the network via the
Cockpit. However, not all parameters are displayed
at the Cockpit (see table below for supported
parameters). In addition, the Cockpit refuses all
low-priority alarm messages but broadcasts all
medium-priority and high-priority alarm messages
to the Infinity network.
Refer to the RS-232 export handbook for
information on which parameters are available for
export protocol. The RS-232 export handbook is
available in English only.
Supported parameters on the Cockpit
The following table lists which MEDIBUS.X
ventilator and anesthesia parameters are displayed
and trended on the Cockpit. Refer to page 403 for
a list of supported ventilators and anesthesia
machines and the compatible software versions.
Parameter
label
Description Unit of
mea-
sure-
ment
Displayed in Ventila-
tion or Anesthesia
Show all page (see
page 413)
Cockpit
Trends page
Available
for display
in parame-
ter field
Yes/No?
Air cons Accumulated air
consumption
L Ventilation Not trended No
Cdyn Dynamic compliance L/bar Ventilation and
Anesthesia
Continuous
trend
No
CO
2 slope Increase of
measured CO2 value
in phase III of the
CO
2 waveform
mmHg/L Ventilation Continuous
trend
No
kPa/L
Vol%L
C20/Cdyn Ratio of compliance
during the last 20%
of inspiration of
dynamic compliance
No units Ventilation Continuous
trend
No
DCO2 CO
2 elimination
coefficient during
HFO
10*mL^2/
s
Ventilation Continuous
trend
No
Des cons Desflurane
consumption
mL Anesthesia Continuous
trend
No
E Elastance mbar/L Anesthesia and
Ventilation
Continuous
trend
No

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals