Troubleshooting
554 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Recording status messages
! %0 Transducer
Unplugged
1), 2)
Parameter value A cable or
transducer has
become
disconnected.
– Reconnect the
cable or
transducer.
– Replace the faulty
part, if the
message persists.
1)
%0 is a placeholder for the parameter label C.O.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
Priority Message Parameter field
Problem Solution
Priority Message Problem Solution
None Primary Recorder Not
Connected
or
Secondary Recorder
Not Connected
A recording was requested
but no recorder is available.
Try again, then contact
DrägerService
®
.
None Primary Recorder Out
of Paper
or
Secondary Recorder
Out of Paper
A recording was requested
but the recorder is out of
paper.
Replace the recorder paper
(see page 494).
None Primary Recorder Door
Open
or
Secondary Recorder
Door Open
The recorder door is open. Close the door of the recorder.
None Primary Recorder
Failure
or
Secondary Recorder
Failure
The recording request was
not accepted due to recorder
hardware failure.
Contact DrägerService
®
.