Maintenance
570 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Inspection / safety checks
Inspection and safety checks of devices must be
performed according to the suggested intervals
specified in the table on page 569.
Scope of inspection/safety checks for
the Cockpit (C500/C700)
The safety checks are no substitute for preventive
maintenance measures (including preventive
replacement of wearing parts) as identified by the
manufacturer.
1 Check accompanying documents:
Instructions for use are available
2 Perform a functional test of the following
features according to the instructions for use:
Verify the LEDs
Perform system tests
3 Check that the device combination is in good
condition:
All labels are complete and legible
There is no visible damage
Fuses which are accessible from the
outside are in compliance with the specified
values
4 Use the instructions for use to check that all
components and accessories needed to use the
product are available.
5 Check the electrical safety requirements
according to IEC62353.
6 Verify that the optical and acoustic alarm
signals function properly.
Scope of inspection/safety checks for
the PS250 / P2500
The safety checks are no substitute for preventive
measures (including preventive replacement of
wearing parts) as identified by the manufacturer.
1 Check accompanying documents:
Instructions for use are available
2 Perform a functional test of the following
features:
Verify the LEDs
Perform system tests
3 Check that the device combination is in good
condition:
All labels are complete and legible.
There is no visible damage.
Fuses which are accessible from the
outside comply with the specified values.
4 Check the electrical safety requirements
according to IEC62353 every two yea rs by
qualified DrägerService personnel.
5 Check the following safety features:
The power LED and the battery indicator
LED function properly.
The C500/C700 are powered correctly.
Check the functional integrity of the Infinity
MCable – Nurse call.
WARNING
Risk of medical device failure
If safety checks are not performed on a
regular basis, the proper operation of the
medical device can be compromised.
Perform safety checks at the indicated
intervals.

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals