Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 397
External device – continuous cardiac output (CCO)
CCO precautions
WARNING
To reduce the risk of patient injury due to
electrical shock, always position the external
device connectivity cable as far from the
patient as possible. Make sure that any cables
or other conducting devices do not come in
contact with the patient. The device
connectivity cable is electrically isolated from
the monitor and any peripheral devices, but
the cable’s enclosure is not electrically
isolated from the peripheral device itself.
WARNING
The Cockpit does not annunciate alarms for
external device parameters.
WARNING
Always refer to the primary data source before
making diagnostic or therapeutic decisions.
Connecting peripheral devices is supported
via the device connectivity option. Only
connect peripheral medical devices to a
patient monitor if those devices comply with
the electrical safety requirements found in IEC
60601-1. Refer to "Safe connection with other
electrical equipment" on page 13 of these
instructions for use for information on how to
connect devices safely.