Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 485
System configuration
Saving profiles
For each patient category, you can create and save
five unique profiles. Included in the five profiles is a
Dräger default profile that cannot be modified. The
following diagram shows the Save Profile page
where you can modify existing profiles and save
new ones.
A Profiles tab
B Save Profile tab
C Adult selection button
D Pediatric selection button
E Neonate selection button
F Profile arrow button
G Description field of selected profile
H Save profile as... button
I Save Profile button (see page 486)
Modifying an existing profile
You can save changes to an existing profile. The
only profile you cannot modify is the Dräger default
profile. The Dräger default profile is activated the
very first time the system is booted up, after a
software upgrade or after factory defaults are
restored.
The adult profiles, pediatric profiles, neonate
profiles pages have identical setup functions. The
only exception is their content which is patient-
category specific. This is important when you are
resetting all profiles to Dräger profiles. Only the
profiles within the currently selected patient-
category are affected.
To save changes to an existing profile
In the following steps, the letters in parentheses
correspond to the diagram for the Save Profile
page.
1 Make sure the Cockpit is in the correct patient
category before configuring the profiles (see
“Selecting the patient category“on page 105).
2 Make the desired changes to the patient
settings.
3 Select the System setup... button on the main
menu bar.
4 Select the Profiles tab (A)
5 Select the Save Profile tab (B). A password
popup appears.
6 Enter the password and select the Enter button.
060
NOTE
Saving a profile for the patient being monitored
might not be applicable to the care unit. To avoid
any issues or the possibility of an automatic
restart, carefully follow the procedure, “To save