BeneVision N Series Patient Monitor Operator’s Manual 10 - 7
10.6.5 Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
This setting is password protected. For more information, see 39.4.6The Other Tab.
The setting of Alarm Delay is not applied to the apnea alarms and the ST alarms. You can set Apnea Delay and
ST Alarm Delay separately.
10.6.5.1 Setting the Apnea Delay Time
To set the apnea delay time, follow this procedure:
1. Access the Setup page in either of the following waysin either of the following ways:
◆ Select the Alarm Setup quick key → select the Setup tab.
◆ Select the Main Menu quick key → from the Alarm column select Setup.
2. Select Apnea Delay to set the apnea delay time.
10.6.6 Adjusting the Alarm Light Brightness
This setting is password protected. For more information, see 39.4.6The Other Tab.
NOTE
• If you set alarm light brightness to Auto, the monitor automatically adjusts the alarm light brightness
according to the ambient light. The stronger the ambient light is, the brighter the alarm light is.
C.O. TB (°C) Adult: (TB - 1)
Pediatric: N/A
N/A Adult: (TB + 1)
Pediatric: N/A
N/A Same as the
measurement range
C.I. N/A N/A N/A N/A N/A
TFC N/A N/A N/A N/A N/A
RM RR (rpm) awRR × 0.5 or
6 (whichever
is greater)
N/A awRR × 1.5 or
30 (whichever
is smaller)
N/A Adult/pediatric: 6 to 55
Neonate: 10 to 90
PEEP
(cmH
2
O)
(PEEP - 5) N/A (PEEP + 5) N/A Same as the
measurement range
PIP
(cmH
2
O)
PIP - 10 N/A PIP + 10 N/A Same as the
measurement range
MVe (L/
min)
MVe - 2 N/A MVe + 2 N/A Same as the
measurement range
SvO
2
SvO
2
(%) SvO
2
- 5 N/A SvO
2
+ 5 N/A Same as the
measurement range
ScvO
2
(%) ScvO
2
- 5 N/A ScvO
2
+ 5 N/A Same as the
measurement range
rSO
2
rSO
2
Same as the
default alarm
limit
Manual
mode: same
as the default
alarm limit
Auto mode:
remains
unchanged
Same as the
default alarm
limit
Manual
mode: same
as the default
alarm limit
Auto mode:
remains
unchanged
15 to 95
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals