BeneVision N Series Patient Monitor Operator’s Manual C - 15
C.1.13 CO
2
Default Settings
C.1.13.1 General Settings
C.1.13.2 Sidestream CO
2
Default Settings
Item Default Setting
EtCO
2
Alarm switch On
High limit Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult and pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO
2
Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
Apnea Delay Adult and pediatric: 20 s
Neonate: 15 s
RR Source Auto
Speed 6.25 mm/s
Scale 50 mmHg
Waveform Type Draw
Item Default Setting
EtO
2
Alarm switch On
High limit 88%
Low limit 18%
Priority Med
Alarm Outputs Off
FiO
2
Alarm switch On
High limit Adult and pediatric: 100%
Neonate: 90%
Low limit 18%
Priority Med
Alarm Outputs Off
BTPS Compensation Off
O
2
Compensation OR: 100%
Other departments: 21%
AG Compensation 0%
N
2
O Compensation 0%
Auto Standby 60 min

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals