EasyManuals Logo
Home>Mindray>Medical Equipment>BeneVision N Series

Mindray BeneVision N Series User Manual

Mindray BeneVision N Series
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #484 background imageLoading...
Page #484 background image
C - 20 BeneVision N Series Patient Monitor Operator’s Manual
C.1.17 BIS Default Settings
C.1.18 NMT Default Settings
Trends EEG Channels All
Parameters SEF
Trend Length 60 min
DSA EEG Channels All
Parameters SEF
Trend Length 20 min
Power Scale 1 - 64 dB
CSA EEG Channels All
Parameters SEF
Trend Length 20 min
Power Scale 1 - 64 dB
CSA Clipping On
Item Default Setting
Item Default Setting
BIS Alarm switch On
High limit 70
Low limit 20
Priority Med
Alarm Outputs Off
Smoothing Rate 15 sec
Display OR: BIS Trend
Other departments: EEG LT/RT
Scale 100 v
Speed 25mm/s
Filter On
Auto Check On
Select Parameter SQI, EMG, SR, SEF
BIS Expand View
Default Setting
EEG Waveforms All
Parameter 1 BIS L
Parameter 2 EMG
Trend Length 60 min
Item Default Setting
Stimulation Current Supra (60 mA)
Pulse Width 200 s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals