24 - 4 BeneVision N Series Patient Monitor Operator’s Manual
• The watertrap collects water drops condensed in the sample line and therefore prevents them from
entering the module. To avoid blocking the airway, empty the watertrap container whenever half
full. Dispose of accumulated fluids in accordance with hospital policy or your local regulations.
• The watertrap has a filter preventing bacterium, water and secretions from entering the module.
After long-term use, dust or other substances may compromise the performance of the filter or even
block the airway. In this case, replace the watertrap. Replacing the watertrap once a month is
recommended.
NOTE
• Do not apply adult watertrap to the neonatal patient. Otherwise, patient injury could result.
• To extend the lifetime of the watertrap and module, disconnect the watertrap from the module and
set the operating mode to Standby when AG monitoring is not required.
24.7 MAC Values
Minimum alveolar concentration (MAC) is the minimum concentration of the agent in the alveoli. It is a basic
index to indicate the depth of anesthesia. The standard ISO 80601-2-55 defines MAC as this: alveolar
concentration of an inhaled anesthetic agent that, in the absence of other anesthetic agents and at equilibrium,
prevents 50% of patients from moving in response to a standard surgical stimulus.
MAC values are listed below:
* indicates 1 MAC nitrous oxide can only be reached in hyperbaric chamber.
NOTE
• The MAC values shown in the table above are those published by the U.S. Food and Drug
Administration for a healthy 40-year-old adult male patient.
• In actual applications, the MAC value may be affected by age, weight and other factors.
The formula to calculate the MAC value is as follows:
Where N is the number of all agents (including N
2
O) that the AG module can measure, EtAgent
i
is the
concentration of each agent, and AgentVol
age
i is the concentration of each agent at 1 MAC with age correction.
The formula for calculating age correction of 1 MAC is:
For example, the Des concentration at 1 MAC of a 60-year old patient is.
The AG module measures 4% of Des, 0.5% of Hal and 50% of N
2
O in the patient’s end-tidal gas:
NOTE
• The formula above is only suitable for patients who are older than one year. If the patient is less than
one year, the system uses one year old to do age correction.
Agent Des Iso Enf Sev Hal
N
2
O
1 MAC 6% 1.15% 1.7% 2.1% 0.77% 105%*
MAC
EtAgent
AgentVol
age
i
-------------------------------------
i0
=
N1
–

=
MAC
age
MAC
40
10
0.00269 age 40
–()
×
–()
×
=
6% 10
0.00269 60 40
–()×–()
×
6% 0.88
×=
MAC
4.0%
6% 0.88×
------------------------
0.5%
0.77% 0.88×
---------------------------------
50%
105% 0.88×
------------------------------++
2.04
==

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals