5 - 4 BeneVision N Series Patient Monitor Operator’s Manual
5.5.2 Transferring Patient Data
To transfer the patient data via N1/T1/MPM, insert the N1/T1/MPM into the module rack or SMR.
â–  If the patient demographics in the monitor are consistent with those of in the N1/T1/MPM, the N1/T1/MPM
automatically uploads the data to the monitor.
â–  If the patient demographics in the monitor are not consistent with those of in the N1/T1/MPM, and Data
Transfer Strategy is set to Always Ask (for more information, see5.5.2Transferring Patient Data), the
monitor prompts the Select Patient menu automatically. In this case, you need to select an operation (see
the following table) according to the actual situation.
NOTE
• If you select Apply Module Settings, the N1/T1/MPM settings can be transferred to the monitor
along with the patient data. For more information, see5.6Exporting Patient Data.
5.6 Exporting Patient Data
To export the data of the current patient and discharged patients, follow this procedure:
1. For N17/N15/N12/N12C, connect the USB drive to the monitor’s USB connector. For N22/N19. connect the
USB drive to the monitor’s MSB connector.
Settings Monitor settings
(Alarm pause, alarm
volume, etc.)
Yes No No No
Parameter settings
(Alarm limits,
measurement setting,
etc.)
Yes Yes Yes Yes
Type of storage Can be stored
in the monitor?
Can be
transferred via
MPM?
Can be
transferred via
T1?
Can be
transferred via
N1?
Operations Operation Description Examples of applications
Continue Patient in Monitor Continue to use the patient data in the
monitor. This deletes all patient data in the
N1/T1/MPM and copies all data in the
monitor to the N1/T1/MPM.
1. Replace N1/T1/MPM during patient
monitoring.
2. After the patient is admitted, connect the
N1/T1/MPM.
Continue Patient in Module Continue to use the patient data in N1/T1/
MPM. The monitor discharges the patient,
and automatically admits a new patient
and copies all data from N1/T1/MPM.
You are monitoring a patient using N1/T1/
MPM, and you need to transfer the patient,
e.g. from a ward (original monitor) to the
operating room (destination monitor).
New Patient Select this option if you will not use either
the information in the monitor or that in
the N1/T1/MPM. This deletes all data in the
monitor and N1/T1/MPM and lets you
admit a new patient on the monitor. In this
case, you need to re-enter the patient
demographics. The monitor will restore the
settings according to the patient category.
Connect the N1/T1/MPM before admitting
a new patient. However, the monitor and/
or N1/T1/MPM have stored the previous
patient’s data and settings.
Same Patient Select this option if the patient information
in the monitor and N1/T1/MPM are
different but you are sure that it is the same
patient. This merges the patient’s trend
data in the monitor and N1/T1/MPM and
copies the settings in N1/T1/MPM to the
monitor as well.
A patient monitored with the N1/T1/MPM
is moved to another department and again
moved back.
However, the patient information stored in
the N1/T1/MPM was altered before
connected to the original monitor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals