EasyManuals Logo
Home>Mindray>Medical Equipment>BeneVision N Series

Mindray BeneVision N Series User Manual

Mindray BeneVision N Series
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #373 background imageLoading...
Page #373 background image
BeneVision N Series Patient Monitor Operator’s Manual 39 - 5
39.4.2 The Pause/Reset Tab
Section Menu Item Default Setting Description
Pause Pause Alarm Pause Selects the pause function.
Alarm Pause: pauses alarms.
Audio Pause: pauses alarm tones.
Pause Time 2 min Selects the alarm pause time.
The alarm pause time can be set to
1 min, 2 min,
3 min, or Permanent.
Pause Priority All Selects alarms of what priority can be paused.
All: pressing the Alarm Pause quick key
pauses all alarms.
Med & Low: pressing the Alarm Pause quick
key pauses alarms of medium and low
priority. The high priority alarms will not be
paused.
Disabled: the Alarm Pause quick key is
disabled.
Pause 5 min Off Selects how long the alarm can be paused if
switched on..
Pause 10 min Off
Pause 15 min Off
Alarm Reset Alarm Light On When Reset
On When Reset: when the alarm system is
reset, the alarm tones of the current alarms
are switched off, but the alarm lamp remains
flashing.
Off When Reset: when the alarm system is
reset, both the alarm tone and alarm lamp of
the current alarms are switched off.
Reminder Tone Alarm Reset
Reminder
On Selects the reminder tone rule when the alarm
volume is set to zero, or the alarm is reset or
switched off.
On: the monitor issues reminder tones at a
designated interval.
Re-alarm: if the alarm condition persists the
acknowledged alarms marked with “√” will be
regenerated after the designated reminder
tone interval.
Off: the monitor does not issue reminder
tones at a designated interval. The
acknowledged alarms marked with “√” will be
silenced.
Alarm Off Reminder On /
Reminder Interval 5 min
10 min: the monitor issues reminder tones
every 10 minutes.
5 min: the monitor issues reminder tones
every five minutes.
3 min: the monitor issues reminder tones
every three minutes.
2 min: the monitor issues reminder tones
every two minutes.
1 min: the monitor issues reminder tones
every one minute.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals