BeneVision N Series Patient Monitor Operator’s Manual D - 1
D Alarm Messages
D.1 Physiological Alarm Messages
This section lists physiological alarms, their default priority, and the actions that can be taken when an alarm
occurs.
D.1.1 General Physiological Alarm Messages
Note: XX represents a measurement or parameter label, such as HR, NIBP, PVCs, RR, SpO2, PR, and so on.
D.1.2 Arrhythmia Alarm Messages
Alarm messages Default priority Cause and solution
XX High Med XX value has risen above the high alarm limit or fallen below the low
alarm limit. Check the patient’s condition and check if the patient
category and alarm limit settings are correct.
XX Low Med
Alarm message Default priority
Asystole High
VFib/VTac High
Vtac High
Vent Brady High
Extreme Tachy High
Extreme Brady High
PVCs/min Med
Pauses/min Med
R on T Med
Bigeminy Med
Trigeminy Med
Tachy Med
Brady Med
Multiform PVC Med
Vent. Rhythm Med
Nonsus. Vtac Med
Run PVCs Low
Pause Low
Couplet Prompt
PVC Prompt
Irr Rhythm Prompt
Pacer Not Pacing Prompt