EasyManuals Logo
Home>Mindray>Medical Equipment>BeneVision N Series

Mindray BeneVision N Series User Manual

Mindray BeneVision N Series
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #480 background imageLoading...
Page #480 background image
C - 16 BeneVision N Series Patient Monitor Operator’s Manual
C.1.13.3 Microstream CO
2
Default Settings
C.1.13.4 Mainstream CO
2
Default Settings
C.1.14 Gas Default Settings
Operating Mode Measure
Item Default Setting
Item Default Setting
BTPS Compen Off
Maximum Hold 20 sec
Auto Standby Off
Operating Mode Measure
Item Default Setting
Maximum Hold 10 sec
O
2
Compensation Off
Balance Gas Room Air
AG Compensation 0%
Operating Mode Measure
Item Default Setting
EtCO
2
Alarm switch On
High limit Adult and pediatric: 50 mmHg
Neonate: 45 mmHg
Low limit Adult and pediatric: 25mmHg
Neonate: 30mmHg
Priority Med
Alarm Outputs Off
FiCO
2
Alarm switch On
High limit 4 mmHg
Priority Med
Alarm Outputs Off
EtO
2
Alarm switch On
High limit 88%
Low limit 18%
Priority Med
Alarm Outputs Off

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals