BeneVision N Series Patient Monitor Operator’s Manual 16 - 3
16.4 Measurement Modes
There are three NIBP measurement modes:
■ Manual: measurement on demand.
■ Auto: repeated measurements at set interval.
■ STAT: continually rapid series of measurements over a five minute period.
■ Sequence: continually automatic measurement at set durations and intervals.
16.5 NIBP Display
The NIBP display shows only numerics.
• If NIBP measurement fails, “XX” is displayed; if NIBP measurement is not taken, “--” is displayed.
• Outlined NIBP numerics indicate that the measurement is old and exceeds the set time. So these
NIBP values are not recommended for reference.
(3) (4) (5)
(10) (9)
(2)
(1)
(6)
(7)
(8)
(11)
(1) Systolic pressure alarm limits (2) NIBP unit: mmHg or kPa
(3)The last NIBP measurement time
(4) Time to the next measurement (for Auto mode and Sequence mode)
(5) Measurement mode: for Auto NIBP, interval is displayed; for Sequence mode, the current phase and interval are
displayed
(6) Diastolic pressure (7) Diastolic pressure alarm limit
(8) Mean pressure alarm limit
(9) Mean pressure (displayed after measurement completed) or cuff pressure (displayed during the measurement)
(10) Systolic pressure (11) Pulse Rate