D - 4 BeneVision N Series Patient Monitor Operator’s Manual
D.2.1 General Technical Alarm Messages
Note: XX represents a measurement or parameter label, such as HR, RR, SpO2, EtCO2, and so on.
D.2.2 ECG Technical Alarm Messages
D
Note: XX represents ECG lead name, for example RL, LL, V, Va, Vb, and so on.
D.2.3 Resp Technical Alarm Messages
Alarm message Default priority
Indication on
alarm reset
Cause and solution
XX Module Error High C XX module does not work properly. Replug the
module, if the alarm persists, contact your service
personnel.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
ECG Noisy Low/Prompt A The ECG signal is noisy. Check for any possible
sources of signal noise around the cable and
electrode, and check the patient for excessive
motion.
ECG Amplitude Too Small Low C The ECG amplitude does not reach the detected
threshold. Check for any possible source of
interference around the cable and electrode.
ECG Lead Off Low B The electrode has become detached from the
patient or the lead wire has become
disconnected from the adapter cable. Check the
connections of the electrodes and leadwires.
ECG XX Lead Off Low B The electrode has become detached from the
patient or the lead wire has become
disconnected from the adapter cable. Check the
connections of the electrodes and leadwires.
ECG Signal Invalid Low A Patient skin impedance is too high. Check ECG
electrode application.
ECG Learning Prompt / ECG learning is manually or automatically
triggered.
Cannot Analyze QT Prompt / /
D12L not available Prompt C The current Va and Vb combination does not
support D12L. Choose an available Va and Vb
combination. For more information, see
11.5Using 6-lead Placement to Derive 12-lead ECG
(D12L)
.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Resp Interference Prompt / The respiration circuit is disturbed. Check for any
possible sources of signal noise.
Electrode Poor Contact Prompt / Check the electrode application. Reposition or
replace the electrodes if necessary.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals