D - 6 BeneVision N Series Patient Monitor Operator’s Manual
D.2.6 NIBP Technical Alarm Messages
D.2.7 IBP Technical Alarm Messages
TempIF Thermometer
Error (for tympanic
thermometer)
High C The tympanic thermometer may fail. Take
temperature measurement again. If the alarm
persists, replace the tympanic thermometer.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Alarm message Default priority
Indication on
alarm reset
Cause and solution
NIBP Cuff Loose Low A There is a leak in the cuff or air tubing. Use a cuff
of correct type based on the patient size. Apply
the cuff and connect the air tubing as instructed
in the manual.
NIBP Cuff or Airway Leak Low A Check the NIBP cuff and pump for leakages.
NIBP Airway Error Low A The air tubing may be occluded. Check the air
tubing for an occlusion or kinking. If the alarm
persists, contact your service personnel.
NIBP Weak Signal Low A The patient’s pulse is weak or the cuff is loose.
Check the patient’s condition and replace the cuff
application site.
NIBP Overrange Low A The measured NIBP value exceeds the module
measurement range. Check the patient’s
condition.
NIBP Excessive Motion Low A Check the patient’s condition and reduce patient
motion.
NIBP Cuff Overpressure Low A The NIBP airway may be occluded. Check the
airway and measure again. If the alarm persists,
contact your service personnel.
NIBP Timeout Low A The measurement time exceeds 120 seconds in
the adult or pediatric mode, or exceeds 90
seconds in the neonatal mode, and the BP value
cannot be obtained. Check the patient’s
condition and NIBP connections, or replace the
cuff and measure again.
NIBP Cuff and Patient
Mismatch
Low A
The cuff type mismatches the patient category.
Verify the patient category or replace the cuff if
necessary. If patient catergory is correct, check
that the tubing is not bent and the airway is not
occluded.
NIBP Airway Leak Low A Airway leakage is found during the NIBP leakage
test. Check the NIBP cuff and pump for leakages.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
XX Sensor Error Med C The IBP sensor fails. Replace the sensor.
XX No Sensor High, Med, or
Low,
configurable
A The IBP patient cable and/or corresponding IBP
sensor is not connected or detached. Check the
cable and sensor connection.
XX No Pulse Low A The catheter may be occluded. Please flush the
catheter.