17 - 4 BeneVision N Series Patient Monitor Operator’s Manual
1. Zero the Codman ICP transducer. For more information, see section 17.4.1Zeroing the Codman ICP
transducer.
2. Disconnect the ICP transducer and ICP adapter cable. Follow the manufacturer’s instructions to apply the
ICP transducer to the patient.
3. Reconnect the ICP transducer and ICP adapter cable.
4. Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP
transducer.
â—† Consistent: select Accept.
â—† Inconsistent: input the zero reference value recorded on the ICP transducer, and select Accept.
If you have to transfer the patient who is taking ICP measurement, check that the target monitor supports the
Codman ICP transducer. For more information, see 17.4.1Zeroing the Codman ICP transducer. If the target
monitor does not support the Codman ICP transducer, do not use it for ICP monitoring.
Follow this procedure to transfer the patient:
1. Disconnect the ICP adapter cable from the measurement module, or remove the module from the monitor.
2. Connect the ICP adapter cable, measurement module, and the target monitor, or insert the measurement
module into the target monitor.
3. Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP
transducer.
â—† Consistent: select Accept.
â—† Inconsistent: input the zero reference value recorded on the ICP transducer, and select Accept.
CAUTION
• If monitors of different brands are used to zero the Codman ICP transducer, the zero reference
values can be different. Use a Mindray monitor to Zero the Codman ICP transducer if you will take
ICP measurement using a Mindray monitor. Otherwise the ICP measurement can be inaccurate.
17.5 IBP Display
The IBP measurement is displayed on the monitor as a waveform and numeric pressures. For arterial pressure,
the IBP numeric area displays systolic pressure, diastolic pressure and mean pressure. For venous pressure, the
IBP numeric area displays only the mean pressure. The figure below shows the waveform and numerics for the
Art pressure.
(1) Pressure label (2) Waveform
(3) Pressure Unit (4) Systolic pressure
(5) Diastolic pressure (6) Mean pressure
(7) PPV measurement
(7) (6)
(1) (2) (3) (4) (5)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals