BeneVision N Series Patient Monitor Operator’s Manual A - 17
PR from IBP Module
A.15.5 Temp Specifications
A.15.5.1 Temp Specifications from the MPM Module and Temp module
A.15.6 Temp Specifications from Genius
TM
2 Tympanic Thermometer
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Standard Meet the standard of ISO 80601-2-56
Technique Thermal resistance
Operating mode Direct mode
Measurement range 0 to 50 °C (32 to 122 °F)
Resolution 0.1°C
Accuracy ±0.1 °C or ±0.2 °F (excluding probe error)
Refreshing rate ≤1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit Range Step
Txx High (xx refers to temperature site) (low limit +1.0) to 50.0 °C
(low limit +2.) to 122.0 °F
0.1 °C
0.1 °F
Txx Low (xx refers to temperature site) 0.1 to (high limit - 1.0) °C
32.2 to (high limit - 2.0) °F
T High 0.1 to 50.0 °C
0.2 to 90.0 °F
Measurement range 33 to 42 °C (91.4 °F to 107.6 °F)
Resolution 0.1°C or 0.1 °F
Calibrated accuracy ±0.1 °C with ambient temperature 25 °C, target temperature 36.7 to 38.9 °C
±0.2 °C with ambient temperature 16 °C, target temperature 33 to 42 °C)
Ambient temperature range 16 to 33 °C (60.8 °F to 91.4 °F)
Response time <2 s
Alarm limit Range Step
TempIF High (low limit +1.0) to 41.9 °C 0.1 °C
TempIF Low 33.1 to (high limit - 1.0) °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals