BeneVision N Series Patient Monitor Operator’s Manual 3 - 11
â–  Alarm Volume: 2
â–  QRS Volume: 1
â–  Key Volume: 0
â–  NIBP End Tone: Off
â–  Stop NIBP: Off
CAUTION
• Verify the night mode settings before entering the night mode. Pay attention to the potential risk if
the setting value is low.
3.7.3.2 Exiting the Night Mode
To cancel the night mode, follow this procedure:
1. Select the Night Mode quick key, or select the Main Menu quick key → from the Display column select
Exit Night Mode.
2. Select Ok.
NOTE
• If you monitor is connected to the CMS, it automatically exits the night mode when being
disconnected from the CMS.
• The monitor resume the previous settings after exiting the night mode.
3.7.4 Standby Mode
You can temperately stops patient monitoring without switching off the monitor by entering the standby mode.
3.7.4.1 Entering the Standby Mode
1. Select the Standby quick key, or select the Main Menu quick key → from the Patient Management
column select Standby.
2. Set Location to define where the patient is when the monitor enters the standby mode.
3. Select Ok.
The monitor behaves as follows after entering the standby mode:
â–  Stops all parameter measurements.
â–  Disables all the alarms and prompt messages, except for the battery low alarm.
â–  Turns screen brightness to the dimmest after entering the standby mode for 30 seconds.
WARNING
• Pay attention to the potential risk of placing the monitor to standby. In the standby mode, the
monitor stops all parameter measurements and disable all the alarm indications, except for the
battery low alarm.
3.7.4.2 Changing the Patient Location at Standby
If you need to change the patient’s location, select Location from the Standby screen.
3.7.4.3 Exiting the Standby Mode
To exit the standby mode, choose any of the following ways:
â–  Select Resume Monitor to exit the standby mode and resume monitoring the current patient.
â–  Select Discharge Patient to discharge the current patient.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N Series and is the answer not in the manual?

Mindray BeneVision N Series Specifications

General IconGeneral
BrandMindray
ModelBeneVision N Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals