GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 3 - Setup Instructions 3-97
3-7-6 Printer Remote Control Selection
To assign a Printer to the remote keys P1, P2 and/or P3:
1.) Press the UTILITIES
key on the keyboard.
2.) In the “Utilities” menu on the left side of the screen select SYSTEM
to invoke the setup desktop.
3.) Select the
P1-P2-P3 page. (If not currently shown, select the [Print] tab).
Configure “Remote” Buttons: Select the desired remote control button.
Select Printer: Select the desired Printer for the selected remote control button.
- If it is desired, check mark “For Reports use Report Printer.”
Report Printer (all Buttons): Select the desired Report Printer from the drop-down menu.
Figure 3-121 P1-P2-P3 page - Print tab
!! NOTICE:
The selected Report Printer is usually used for printing reports and images from Sonoview.
Report Printer
Select Printer
Configure Buttons