GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-45
8-19-1 Replacement of the Probe Holder Kit
8-19-1-1 Manpower
One person, 5 minutes
8-19-1-2 Tools
none
8-19-1-3 Preparations
none
8-19-1-4 Probe Holder Kit - Removal Procedure
1.) Pinch together the Probe Holders.
2.) Remove the Probe Holders.
8-19-1-5 Probe Holder Kit - Installation Procedure
1.) Place the new Probe Holder.
Figure 8-60 remove the Probe Holder