GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 3 - Setup Instructions 3-3
3-2-3 Safety Reminders
!! DANGER:
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT
TOUCH THE UNIT!
!! CAUTION:
Two people are required to unpack the optional Cart because of its weight.
Two people are required whenever a part weighing 16kg (35 lbs.) or more must be lifted.
!! CAUTION:
If the unit is very cold or hot, do NOT turn on its power until it has had sufficient time to
acclimate to its operating environment.
!! CAUTION:
To prevent electrical shock, connect the unit to a properly grounded power outlet.
Do NOT use a three to two prong adapter. This defeats safety grounding.
!! CAUTION:
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
!! CAUTION:
Do not operate this unit unless all board covers and frame panels are securely in place, to
ensure optimal system performance and cooling.
(When covers are removed, EMI may be present).
!! CAUTION:
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the
Voluson i / Voluson e.
Keep manuals near the unit for reference.
!! CAUTION:
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Voluson i / Voluson e portable
ultrasound system is within FDA limitations, avoid unnecessary exposure.
Ultrasound energy can produce heat and mechanical damage.
Figure 3-1 Environmental Labels