IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-3
Section 8-2
System Software - Installation/Upgrade Procedure
8-2-1 FMI from DVD
8-2-1-1 Introduction
8-2-1-2 Manpower
One person ~ 1 hour (depends on contents of System DVD, peripherals, etc.)
8-2-1-3 Tools
System DVD
8-2-1-4 Preparations
Before performing the Software Update/Upgrade:
A.) make sure that all system functions are working correct
B.) check the current Ultrasound Application Software version and the installed Options
C.) if the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 8.2.0 or higher, calculate new software specific “Permanent key” in OKOS.
NOTE: It is NOT necessary to save Full System Configuration (Full Backup) prior to the upgrade.
All existing User Programs, 3D/4D Programs and Auto Text settings remain untouched!
!! NOTICE:
To update the Systems C:\ image via the FMI FROM DVD button in the System Setup SERVICE page,
see: Section 8-2-1-5 "System Software - Installation Procedure (FMI from DVD)" on page 8-7.
The Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS Patches, etc.) depend on the contents of the System DVD being used.
!! NOTICE:
f the currently installed software has to be upgraded to a newer version, and the system DVD being
used is SW 8.2.0 or higher, a new software specific “Permanent key” is required.
(Calculated in OKOS
Option Key Order System; http://3.187.187.9/OKOS.
If you do not have permission to access OKOS, contact the On Line Center to get the necessary key.