GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
4-32 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-3 Save Full Backup (Presets, Configurations & Application Settings)
A full backup always contains the following data
Patient demographic and exam data (database containing the patient data and measurements)
SonoView image data (NOT available when saving to the internal hard disk or DVD/CD)
User Settings (databases and files containing gray curves and the user settings.)
Image transfer settings (DICOM® settings e.g., DICOM® servers, AE Title, Station Name, etc.)
Measure Setup Settings (user specific measure settings)
Machine settings (general settings such as language, time/date format and the enabled options)
Windows Network Settings (network settings including the computer name)
Serviceplatform (state of the Service Software)
VP (additional system data)
NOTE: Always “Full Backup” any presets, configurations and application settings to HDD and/or DVD before
upgrading the software and/or application settings. This ensures that if the presets need to be reloaded,
will be the same ones the customer was using prior to service.
1.) Press the UTILITIES
key on the keyboard. The menu area changes to the Utilities menu.
2.) Select the SYSTEM
item from the menu area to activate the setup desktop screen.
3.) Select the
BACKUP page and click the SAVE button of the “Full Backup” group.
4.) Choose the Destination.
5.) Enter the description of the full backup.
6.) If desired and possible (Network Drive and Other drive only), activate “Include Images.
NOTE: The “Include Images” option may result in a large amount of data: up to 70 Gigabytes!
7.) Select the NEXT
button. To start the backup process click YES.
After copying the data, the Voluson i / Voluson e reboots and the application starts again.
!! CAUTION:
It is recommended to create a full backup of the settings once a week.
Figure 4-22 Full Backup Save
choose destination
enter description
click Next

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson i BT06 and is the answer not in the manual?

GE Voluson i BT06 Specifications

General IconGeneral
BrandGE
ModelVoluson i BT06
CategoryMedical Equipment
LanguageEnglish

Related product manuals