IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-21
Section 8-3
User Settings Only (Application Settings) Loading Procedure
8-3-1 Introduction
The User Settings contains:
• User Programs
• Auto Text
• 3D/4D Programs
8-3-2 Loading Procedure
see: Section 4-5-2 "Load User Settings Only (Application Settings)" on page 4-30
Section 8-4
Full Backup (Presets, Configurations & Appl. Settings) Loading Procedure
8-4-1 Introduction
The Full Backup contains following data:
• Patient demographic and exam data (database containing the patient data and measurements)
• SonoView image data (NOT available when saving to the internal hard disk or DVD/CD)
• User Settings (databases and files containing gray curves and the user settings.)
• Image transfer settings (DICOM® settings e.g., DICOM® servers, AE Title, Station Name, etc.)
• Measure Setup Settings (user specific measure settings)
• Machine settings (general settings such as language, time/date format and the enabled options)
• Windows Network Settings (network settings including the computer name)
• Serviceplatform (state of the Service Software)
• VP (additional system data)
8-4-2 Loading Procedure
see: Section 4-5-4 "Load Full Backup (Presets, Configurations & Application Settings)" on page 4-34