IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-19
10-7-7-3 Data Sheet for Enclosure/Chassis Leakage Current
Table 10-14 shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by system.
Record all data in the Electrical safety tests log.
Table 10-14 Typical Data Format for Recording Chassis Source Leakage
Unit Under Test: ________________________ Date of Test: _____________________
Test Conditions Measurement/Test Point Location
System Power Grounding/PE
Potential
equilibrium
connector
Monitor
housing
Probe
connector
off closed
off open
on closed
on open