GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-83
8-20-22 Replacement of the Power Supply Patient Monitor
8-20-22-1 Manpower
One person, 5 minutes
8-20-22-2 Tools
none
8-20-22-3 Preparations
1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
8-20-22-4 Power Supply Patient Monitor - Removal Procedure
1.) Disconnect the Power Supply on the rear side of the Voluson Station.
8-20-22-5 Power Supply Patient Monitor - Installation Procedure
1.) Connect the new Power Supply.
Figure 8-104 disconnect the Power Supply