IRECTION KTI106052, REVISION 10 SERVICE MANUAL
5-58 Section 5-14 - Boot Screen Functions
5-14-1-3 Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during “FMI from DVD”.
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-32 on page 5-57), press the PG DN
[ Arrow down] key on the keyboard until the Rollback item is highlighted, then press ENTER
.
3.) When the following WARNING message appears, press the
[Arrow left] button to highlight OK
and then press ENTER
.
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
5-14-1-4 Memtest
Starts the PC Memory Test.
Operation see: Section 5-14-2 "Memory Check in LINUX" on page 5-59.
Figure 5-33 Warning message
select [OK]