GE VOLUSON
i / VOLUSON
e
D
IRECTION KTI106052, REVISION 10 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-55
8-19-11 Replacement of the Probe Holder for Endocavity probes
8-19-11-1 Manpower
One person, 5 minutes
8-19-11-2 Tools
Phillips screwdriver 1
8-19-11-3 Preparations
none
8-19-11-4 Probe Holder- Removal Procedure
1.) Screw off the 2 screws and carefully remove the Probe Holder.
8-19-11-5 Probe Holder- Installation Procedure
1.) Place the new Probe Holder and fasten it with the screws as shown in Figure 8-75 on page 8-55.
Figure 8-75 screw off the screws