GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 4 Functional Checks 4-3
4-3-1-2 Back-end Processor Boot Up
Press the Standby switch on the Control Panel once.
When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the
software code is distributed to initiate the scanner.
No status messages are displayed during this process.
4-3-2 Power Off / Shutdown
NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-2-1 Back End Processor Power Off
1.) Press the Standby switch on the Control Panel once to display the SYSTEM - EXIT menu.
2.) Select Shutdown from the SYSTEM - EXIT menu.
3.) Wait for the system to properly close all files and shutdown.
Figure 4-2 Standby Switch for Back-end Processor Boot Up
Figure 4-3 System Exit Menu for Back-end Processor Power Down
Standby Switch