7-78 Section 7-13 - Troubleshooting TiPs
7-13-7-2 No Remote Control for Peripherals
Check cable connections.
Check configuration of the device.
Clean User Defs before reloading software in case there is a corrupt Connectivity.res file. Be sure
to do an Alt-D to capture logs before performing a Clean User Defs so that if the presets prove
corrupt the logs can be sent to OLC for Engineering to review.
Reload software.
Replace the device.
7-13-7-3 No Video on LCD Display
If the video is too dim, has dimmed areas or there is no video at all, replace the HV LCD inverter.
If the HV LCD inverter is burnt, there is a possibility that the Digital Video Card on the PC also got
damaged; in that case you also have to order the Back End Processor.
Replace upper panel assembly.
7-13-7-4 Wrong Key Activated on the Touch Panel
Perform Touch Screen Verification Utility, check first before calibrating. See section 7-7-4-10
"Calibration Utilities" on page 7-34. Access the Service interface (Click on the Wrench) and go to
Diagnostics > Acquisition diagnostics > Diagnostics groups > Calibration Utilities > Start Touch
Screen Verification.
Calibrate touch panel. See section 7-7-4-10 "Calibration Utilities" on page 7-34. Access the Service
interface (Click on the Wrench) and go to Diagnostics > Acquisition diagnostics > Diagnostics
groups > calibration utilities > Touch panel calibration
Replace Upper Panel Assembly.
7-13-7-5 Touch Panel Not Responding
Perform Touch Screen Verification Utility, check first before calibrating. See section 7-7-4-10
"Calibration Utilities" on page 7-34. Access the Service interface (Click on the Wrench) and go to
Diagnostics > Acquisition Diagnostics > Diagnostics Groups > Calibration Utilities > Start Touch
Screen Verification.
Calibrate touch panel. See section 7-7-4-10 "Calibration Utilities" on page 7-34. Access the Service
interface (Click on the Wrench) and go to Diagnostics > Acquisition Diagnostics > Diagnostics
Groups > Calibration utilities > Touch Panel Calibration.
Reload software.
Replace Upper Panel Assembly.
7-13-8 Probes
7-13-8-1 Probe Recognition
Check all the probes on all the ports. Remove all the probes, and then check each probe singly in
every port.
- If the problems persist with all the probes, replace the XDIF board.
- If only one probe fails to be recognized, clean User Defs to check if the problem is caused by
corrupted presets. If the problem is not resolved, replace the probe.

Table of Contents

Other manuals for GE LOGIQ 9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 9 and is the answer not in the manual?

GE LOGIQ 9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 9
CategoryMedical Equipment
LanguageEnglish

Related product manuals