Chapter 7 Diagnostics/Troubleshooting 7-77
7-13-6-3 Drive Failures (CD or MOD)
Check that the media (disk capacity and speed) is supported.
- For the CD, there are 3 different speeds. Drives are listed like this: 24x10x40x. The first
number (24x) indicates what speed the drive can write to a CD-R disc (write speed only). The
second number (10x) indicates what speed the drive can write to a CD-W (write & re-writes
many times). The last number (40x) indicates how fast the drive can read a CD. You only
need to check the first number (typically 24 or 16) since we only use CD-R discs. The capacity
is normally 700MB.
- For the MOD, the size is 3.5 inches, 230, 540 or 640 MB. The 230 MB disc is the most
commonly used.
Check Windows for driver issues. Go to Control Panel > Administrative tools> Computer
management> Device Manager.
Re-install drivers if necessary.
If the CD drive is having problems, replace BEP.
If MOD drive is having problems, try reloading software.
If the problem persists, replace BEP.
7-13-6-4 Video Problems on the Main Monitor
If the problem appears on the main monitor and it involves the whole screen, not only the image
area, check your peripherals.
- If present on the peripherals, replace BEP.
- If not present on the peripherals, move the connector D4 (SVGA out to the top monitor) on the
BEP rear panel to the C13. That connects the monitor directly to the SVGA port. If the problem
is still present, replace Monitor. If not, replace BEP.
7-13-6-5 Simulator Mode
Check the connection of the PC2IP cable on both the SCB board and BEP.
Check the part number on the SCB board. It should be 2260214-2 or 2260214-3. The revisions with
earlier (lower) numbers had issues with the PCI connector. The connector on the board had
intermittent contact on the 2260214 boards. If the SCB has an older revision, order FMI 70168
Check FEPS. You can visually check the LED status on the FEPS. If any of the voltages is missing,
replace FEPS.
7-13-7 Operator Panel
7-13-7-1 No Audio
Check volume settings in the application and also in Windows.
Use headphones, (the type that you use on a personal CD player or a laptop) to test the audio
output directly from the back of the BEP.
- If no Audio is present on the BEP, Reload software. If the problem persists, replace the BEP.
- If Audio is present, follow the audio signal to the VCR or the OP Panel (the audio amplifier is
located in the upper OP panel). The Audio output from the BEP goes to the Internal I/O and
from there to the VCR and OP Assembly (use an adapter to plug the headphones to the RCA
outputs of the Internal I/O). See Section 5-5 "Operator Console Control PWA" on page 5-31.
If Audio outputs are working, replace Upper OP Assembly. Otherwise replace Internal I/O
board.
- Always measure the speakers’ impedance; it should measure approximately 7 ohms. If
Speakers are bad, it is possible that the amplifier on the Upper OP Assembly might be
defective, too.

Table of Contents

Other manuals for GE LOGIQ 9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 9 and is the answer not in the manual?

GE LOGIQ 9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 9
CategoryMedical Equipment
LanguageEnglish

Related product manuals