GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
8-10 Section 8-2 - Plastic Covers and Bumpers
8-2-6 Lower Rear Cover Replacement Procedure
8-2-6-1 Manpower
One person, 15 minutes
8-2-6-2 Tools
Phillips Screwdriver
8-2-6-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Remove both left- and right-side covers as shown in Section 8-2-2 on page 8-5.
3.) Remove the upper rear cover as shown in Section 8-2-3 on page 8-7.
8-2-6-4 Lower Rear Removal Procedure
1.) Unscrew the six Phillips screws that fasten the lower rear cover to the chassis shown in Figure 8-6.
8-2-6-5 Lower Rear Cover Installation Procedure
(click here to view the cover installation video)
1.) Align the lower rear cover with the chassis and fasten it with the six Phillips screws.
2.) Install both left- and right-side covers as shown in Section 8-2-2 on page 8-5.
3.) Install the upper rear covers shown in Section 8-2-3 on page 8-7.
4.) If required, Power On/Boot-up the system as described in Section 4-3-1 on page 4-2.
Select the movie camera icon to view the mpeg video of the Cover
removal and replacement procedure.
Figure 8-6 Lower Rear Cover with Bumper
Screws
Lower Rear Cover
Screws
Magnetic lock
5 - 6 washers
3 - 4 washers

Table of Contents