8-102 Section 8-23 - Console Lock Assembly Replacement Procedure
Section 8-23
Console Lock Assembly Replacement Procedure
8-23-1 Manpower
One person, 1 hour
8-23-2 Tools
Hexagon key 2,5 and 3 mm
Phillips screwdriver size 2
Socket 8 mm
8-23-3 Preparations
1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
2.) Remove Operator Panel.
3.) Remove XY- Handle as described in X/Y Release Handle Replacement on page 8-101.
Figure 8-113 Console Lock Assembly
5mm screws
Spring
5 mm screws
3mm screws