GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
4-12 Section 4-3 - General Procedure
4-3-7-3 Deactivating the probe
When deactivating the probe, the probe is automatically placed in standby mode.
1.) Press the Freeze key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for
complete probe cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
4-3-7-4 Disconnecting the probe
Probes can be disconnected at any time. However, the probe should not be selected as the active
probe.
1.) Move the probe locking handle counterclockwise. Pull the probe and connector straight out of
the probe port.
2.) Carefully slide the probe and connector away from the probe port and around the right side of
the keyboard. Ensure the cable is free.
3.) Be sure that the probe head is clean before placing the probe in its storage box.
4-3-8 Using Cine
4-3-8-1 Activating CINE
Press FREEZE
, then roll the TRACKBALL to activate CINE. To start CINE Loop playback, press Run/
Stop. To stop CINE Loop playback. press Run/Stop.
4-3-8-2 Omitting Images
TRACKBALL
to the frame you want to delete and press OMIT IMAGE.
4-3-8-3 Restoring Images
To undo deleting an image from the CINE Loop, press RESTORE IMAGE
.
4-3-8-4 Quickly Move to Start/End Frame
Press FIRST
to move to the first CINE frame; press LAST to move to the last CINE frame.
4-3-8-5 Start Frame/End Frame
Turn the START FRAME
dial to the left to move to the beginning of the CINE Loop. Turn the dial to the
right to move forward through the CINE Loop.
Turn the END FRAME
dial to the right to move to the end of the CINE Loop. Turn the dial to the left to
move backward through the CINE Loop.

Table of Contents

Other manuals for GE LOGIQ 9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 9 and is the answer not in the manual?

GE LOGIQ 9 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 9
CategoryMedical Equipment
LanguageEnglish

Related product manuals