GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
8-186 Section 8-50 - Wireless Voice Scan Option
8-50-4 Transmitter Setup
8-50-4-1 Overview
To use wireless Voice Recognition, your microphone will be connected to a Transmitter which talks to
a Receiver installed in the LOGIQ™ 9. If you have more than one LOGIQ™ 9 with wireless voice
recogintion in your location, each LOGIQ™ 9’s receiver must be set to a unique frequency per 8-50-3
"Receiver Setup" on page 8-182.
Transmitters must then be set to the same frequency as the receiver to talk to the LOGIQ 9. The
Transmitters have 4 presets (which are called channels) so that you can conviently use one transmitter
to talk to 4 different LOGIQ™ 9 receivers.
However, only one transmitter can be used at a time. Do not have multiple transmitters powered on
and tuned to the same frequency and expect to have them both communicate with the same
LOGIQ™ 9. Neither transmitter will work reliably in this situation.