8-148 Section 8-36 - Rear Caster Replacement Procedure
8-36-4 Rear Caster Removal Procedure
1.) Turn Front Casters as shown in Figure 8-154.
2.) Activate Direction Lock.
3.) Put the bevel-edged board on the floor.
4.) Pull the system backwards up the board incline. This will lift the Rear Wheel on the opposite side
of the System from the floor.
5.) Activate the brakes on all Casters, except the one you intend to replace.
6.) Turn the Rear Caster that stands on the Bevel Edged Board in the direction as shown in Figure 8-
155.
WARNINGWARNING
Prior to elevating scanner, verify that the floating keyboard is locked in its lowest
position.
WARNINGWARNING
Use extreme caution as long as LOGIQ™ 9 is un-stable, not resting on all four Casters.
Figure 8-154 Pull system backwards up the board incline