GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
6-2 Section 6-3 - Monitor
Section 6-3
Monitor
6-3-1 Monitor Adjustment
Contrast, brightness and degauss are adjusted using the three (3) keys located on the front of the
monitor just below the screen.
By pressing either the increase button or decrease button the degauss menu appears. Pressing the
menu button will degauss the monitor. If you do not wish to degauss the monitor, again press either the
decrease button or increase button and the monitor will return to the default screen.
To calibrate brightness and contrast, go to Utility>Test Patterns>Calibration. The screen display will
change as shown in Figure 6-2 on page 6-3 .
Figure 6-1 Monitor Adjustment Switches
DECREASE INCREASEMENU