GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
8-146 Section 8-35 - Verify and Update Vital Product Data
8-35-3-5 Acquisition Diagnostics - Write New VPD Data to EEPROM
Once the *.txt files have been updated with the appropriate information, go back to the Netscape
Browser and navigate to the following menu location:
1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware
Information Utilities>Acquisition Hardware Information Write Utilities.
2.) Separately select the Equalization circuit board and Backplane menu items to write the new
information to the EEPROMs. You can check data by reading back the files as performed in
Section 8-35-3-3 on page 8-138.
Figure 8-152 Acquisition Hardware Information Write Utilities