GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARY MANUAL
Chapter 5 Components and Functions (Theory) 5-37
Section 5-7
Internal I/O
The Internal I/O is the interface between the Card Cage, PC Module, and the rest of the system,
including the Top Console and the Peripherals. Table 9-16 on page 26 and Table 9-18 on page 33 have
information on cable call-outs.
Figure 5-26 Internal I/O Basic Interconnect
INTERNAL I/O
PC2IO
PCVIC
TV video,
SGVA
& I2C
Power
SVGA BD.
PC2IP
UPS
PAT I/O
RS232
Power
Power
Serial data, I2C
RS232, I2C
& ctrl. signals
EXTERNAL I/O
PERIPHERALS
(VCR & printers)
AC POWER
& FANS
3D &
ROTATION
ADAPTER
FRONT-END
PROCESSING
MODULE
SVGA
MONITOR
TV - RGBS
TV video, audio,
RS232, ctrl. signals
Ctrl. signals
Ctrl. signals
Ctrl. signals
audio & TV video
Power
TV video, SVGA, RS232 &
I2C
Power
RS232, USB, I2C, Ethernet,
SVGA, TV video
& ctrl signals
Power
TOP CONSOLE
USB, audio, ctrl. signals
Power
SVGASVGA
BACK-END PROCESSING MODULE
Power
Ethernet, USB,
COM1/2 & audio

Table of Contents