Receiver Mode
R&S
®
ESR
185User Manual 1175.7068.02 ─ 12
●
CISPR only
Gaussian filter with a 6 bandwidth. 6 dB bandwidths that comply with CISPR stand-
ards are available.
●
MIL Std only
Gaussian filter with a 6 bandwidth. 6 dB bandwidths that comply with military
standards are available.
●
3 dB Bandwidth
Gaussian filter with a 3 dB bandwidth.
6 dB bandwidths correspond approximately to the pulse bandwidth..
3 dB bandwidths correspond approximately to the noise bandwidth.
Remote command:
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE on page 671
IF Analysis RBW
Selects the resolution bandwidth for IF spectrum analysis
Available for IF Analysis (firmware application R&S ESR-K56).
Remote command:
[SENSe:]BANDwidth:IF on page 670
5.3.4 Configuring the Scan
The following settings for configuring the scan are available in the [Sweep] menu.
Functions to configure measurements described described elsewhere:
●
"Fixed Frequency" on page 189
Run Continuous / Run Single......................................................................................185
Edit Scan Table........................................................................................................... 185
Freq Axis (Lin Log)......................................................................................................185
Run Continuous / Run Single
Initiates a continuous or single measurement.
See also Chapter 5.1.7.1, "Running Scans and Measurements", on page 156.
Remote command:
Selecting single and continuous measurements:
INITiate<n>:CONTinuous on page 641
Initiating a measurement:
INITiate<n>[:IMMediate] on page 641
Edit Scan Table
Opens a dialog box to create or edit a scan table.
For more information see Chapter 5.1.3.3, "The Scan Table", on page 146
Freq Axis (Lin Log)
Switches between linear and logarithmic frequency axis.