I/Q Analyzer
R&S
®
ESR
469User Manual 1175.7068.02 ─ 12
The setting 75 Ω should be selected if the 50 Ω input impedance is transformed to a
higher impedance using a 75 Ω adapter of the RAZ type (= 25 Ω in series to the input
impedance of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω/50 Ω).
All levels specified in this Operating Manual refer to the default setting of the instru-
ment (50 Ω).
Remote command:
INPut:IMPedance on page 670
7.3 Softkeys of the Marker To Menu in I/Q Analyzer Mode
In I/Q Analyzer mode, The "Marker To" menu is identical to the one in Spectrum mode
(see "Softkeys of the Marker To Menu" on page 411). For the "Real/Imag (I/Q)" display
mode, however, an additional function is available.
Search Settings
Opens a dialog box to define which data is used for marker search functions.
Note: The search settings apply to all markers, not only the currently selected one.
"Search Real"
Marker search functions are performed on the real trace of the I/Q
measurement.
"Search Imag"
Marker search functions are performed on the imaginary trace of the
I/Q measurement.
"Search Mag-
nitude"
Marker search functions are performed on the magnitude of the I and
Q data.
Remote command:
CALCulate<n>:MARKer<m>:SEARch on page 888
7.4 Softkeys of the Trigger Menu in I/Q Analyzer Mode
In I/Q Analyzer mode, the "Trigger" menu, which is displayed when you select the
[TRIG] key, contains the following functions:
Trg/Gate Source..........................................................................................................470
Free Run.......................................................................................................470
External.........................................................................................................470
Video.............................................................................................................470
RF Power......................................................................................................470
IF Power/BB Power.......................................................................................471
Time.............................................................................................................. 472
Trigger Level............................................................................................................... 472
Trigger Polarity............................................................................................................472
Trigger Offset.............................................................................................................. 472
Softkeys of the Trigger Menu in I/Q Analyzer Mode