Receiver Mode
R&S
®
ESR
187User Manual 1175.7068.02 ─ 12
5.3.6 Controlling Inputs and Outputs
The input/output settings are available in the [Input/Output ] menu.
For information on the LISN control see Chapter 5.3.7.6, "LISN Settings",
on page 196.
Input (AC/DC)..............................................................................................................187
Input (1 2)....................................................................................................................187
User Port.....................................................................................................................187
Input (AC/DC)
Toggles the RF input of the R&S ESR between AC and DC coupling.
Remote command:
INPut:COUPling on page 672
Input (1 2)
Selects the RF input the signal is applied to.
The first RF Input supports a frequency range from 9 kHz to f
max
and an attenuation
range from 0 dB to 75 dB. The second RF Input is a pulse-resistant input. It supports a
frequency range from 9 kHz to 1 GHz and an attenuation range from 10 dB to 75 dB.
Attenuation levels smaller than 10 dB are not possible at RF Input 2.
With option R&S ESR-B29, the minimum frequency is extended to 20 Hz at both RF
inputs.
Remote command:
INPut:TYPE on page 672
User Port
The user port (AUX port) can transmit bit patterns in two directions, depending on the
actual selected signal direction. The user port is available on the rear panel when you
equip the R&S ESR with option R&S ESR-B5.
●
"Direction" = "In" (read only)
Opens a dialog box. Select the "Read" button to read out the user port configura-
tion. The value is displayed in the dialog box.
●
"Direction" = "Out"
Select the required bit pattern by changing the state of the individual ports ("Port
<x>").
See "AUX PORT" on page 38 for information about the pin assignment of the con-
nector.
See OUTPut:UPORt[:VALue] on page 933 for information about the pin to bit
assignment.
Remote command:
State: INPut<ip>:UPORt:STATe on page 931
State: OUTPut:UPORt:STATe on page 933
Bit pattern (query): INPut<ip>:UPORt[:VALue] on page 932
Bit pattern (setting): OUTPut:UPORt[:VALue] on page 933
Common Measurement Settings